Feature

Rapid test
Testing <2mins with result

Wide Range
Maximum measurement range
in same function POCT meter

Accurate
Same level of accuracy with Biochemical Analyzer
The Triglycerides Test Strip(Dry Chemistry) work with the Analysis Meter to measure the Triglycerides (TG) concentration in whole blood, plasma and serum.
For professional use only.
MEASUREMENT RANGE
Triglycerides: 0.51-7.34 mmol/L (45-650 mg/dL)
Note:for triglycerides, 1mmol/L=88.6mg/dL.
Results below the ranges will show “Lo”, and results above the ranges will show “HI”.
PRINCIPLE
The Triglycerides Test Strip ( Dry Chemistry ) use a timed-endpoint method to measure the Triglycerides(TG) concentrations in whole blood,serum or plasma. The system monitors the change in absorbance at 620 nm at a fixed-time interval. The change in absorbance is directly proportional to the concentration of lipid in the specimen.
TG: Triglycerides in the specimen are hydrolyzed to glycerol and free fatty acids by the action of lipase. A sequence of three coupled enzymatic steps using glycerol kinase (GK), glycerophosphate oxidase (GPO), and peroxidase (POD) causes the oxidative coupling of 4-aminoantipyrine to form a blue dye.
Reference values are listed in the chart below:
Desirable: <1.7 mmol/L (150 mg/dL)
Borderline High: 1.7 -2.3 mmol/L (150-200mg/dL)
High: >2.3 mmol/L (200 mg/dL)
Reference ranges may vary between laboratories. Every laboratory should establish its own reference range as needed.
Blood lipid levels will have big physiological fluctuations depending on food consumed or exercise.
Accurate
Blood Lipid Analysis Items
Triglycerides (TG)

Procedure

Packing