Hemoglobin Test Strip (BS-101)

introduction

Modle

BS-101

Analysis Items

hemoglobin, Hb, hematocrit, HCT

Specimen

Whole blood(peripheral and venous human whole blood), plasma and serum

Features

Wide Test Range, Accurate, Quality Assurance, Mirco Sepcimen

Package

25T/Box;50T/Box;

Feature

Multi-function

hemoglobin (Hb); Hematocrit (HCT)

Wide Range

Maximum measurement range
in same function POCT meter

Accurate

Same level of accuracy with Biochemical Analyzer

The Hemoglobin Testing System is intended for the quantitative determination of hemoglobin (Hb) and calculated hematocrit (Hct) in capillary and venous human whole blood. The easy to operate system consists of a portable meter that analyzes the intensity and color of light reflected from the reagent area of a test strip, ensuring quick and accurate results.

The Hemoglobin Testing System provides results in less than 15 seconds and requires only a single drop of whole blood. The meter can store up to 1,000 results and records can be transferred to a computer for further analysis using the USB port. For Model BC-102,the records can also be transferred to a mobile terminal by bluetooth. The meter can be operated by 3 AAA (1.5V) batteries.

To ensure accurate results:

·Read instructions and complete any necessary training before use.

·Use the code chip that accompanies each box of test strips.

·Use only BioAid® Hemoglobin Test Strips with the BioAid® Hemoglobin Meter.

·For in vitro diagnostic use only.

·For professional use only.

·Test only whole blood samples. EDTA or heparin anticoagulants can be used.

·Keep out of reach of children.

Note: Throughout this user guide, meter parts or functions will appear in bold. Items appearing on displays are identified in bold italics.

PRINCIPLE

Erythrocytes in the specimen are lysed to release hemoglobin. The hemoglobin is converted to methemoglobin. The intensity of the color produced from this reaction is proportional to the hemoglobin concentration. Reference values are listed in the chart below:

Men: 13.0-17.0 g/dL(130-170 g/L, 8.1-10.5 mmol/L)

Women: 12.0-15.0 g/dL(120-150 g/L, 7.4-9.3 mmol/L)

Children: 11.0-14.0 g/dL(110-140 g/L, 6.8-8.7 mmol/L)

Specfication

  • hemoglobin (Hb)

  • Hematocrit (HCT)

FeatureSpecifications
MethodologyReflectance Photometer
Test Timehemoglobin (Hb)|Hematocrit (HCT)≤15 seconds.
Measurement Rangehemoglobin: 4.5-25.6 g/dL (45-256g/L, 2.8-15.9mmol/L)
SpecimenWhole blood(peripheral and venous human whole blood), plasma and serum
Specimen Volume10 μL
Units of Measurementmg/dL, g/dL, mmol/L
Meter storage and transportationt conditions0 - 55°C; ≤90% RH
System Operating Conditions10 - 35°C; ≤80% RH; indoor only

Accurate

Blood hemoglobin Analysis Items

Hemoglobin (hb)

Procedure

Packing

Contents

  • 2 Vials of Hemoglobin Test Strip

  • 1*Code Chip

  • 2* Capillary transfer tube

  • 1*Package Insert

  • 1* Insert of Capillary tranfer tube

Matched Analyzer

related packages